Detail Package

Procurement of USG and Centrifuge

Project
SOPHI
Project Description
Strengthening of Primary Healthcare in Indonesia (SOPHI): a joint co-financing between WB, AIIB and ADB for the primary care component.
Package Name
Procurement of USG and Centrifuge
Loan No.
IBRD (9626-ID), AIIB (L0787A), ADB (4430-INO)
Activity Reference No.
ID-PMU SOPHI-513307-GO-RFB
Market Approach
National Open
Procurement Method
Request for Bids (RFB) - 1 Stage 1 Envelope (Lowest Evaluated Bid)
Estimated Total Cost
Rp127,770,242,000.00
Currency
Indonesian Rupiah
Stages
Technical Evaluation
Bid Submission Deadline
31 Jul 2026

Package Lot

Name Of Lots Units Amount
Lot 1: USG (2D) 536 Rp62,427,920,000.00
Lot 2: Centrifuge 2133 Rp65,342,322,000.00

Documentation

Published Date Document Name
2026-07-03 Specific Procurement Notice - Procurement of USG and Centrifuge
2026-07-03 Bidding Document - Procurement of USG and Centrifuge
2026-07-03 Technical Specification - Procurement of USG and Centrifuge
2026-09-02 Minutes of Prebid Meeting - Procurement of USG and Centrifuge
2026-09-18 1st Responses to Clarification of Bidding Document - Procurement of USG and Centrifuge

Responses to Clarification of Bidding Document (Procurement of USG and Centrifuge)

No. Date of Received Question Responses Part (RFB) Cluster Sub Cluster
1 2026-07-18 "1, Row 43, PWD and continuous wave (CW) sweep speed

Required



The tender specifications require a 2D ultrasound system for OB/GYN and general examinations, specifying only convex and linear probes; however, Row 43 requires the provision of sweep speed specifications for CW (Continuous Wave) mode. Additionally, we note that Rows 40, 41, 42, and 45 all request parameters related to PWD (Pulsed Wave Doppler).



Ultrasound systems operating in CW mode typically utilize dedicated CW Doppler ""pencil"" probes or phased array probes with CW capabilities, which are suitable for measuring extremely high blood flow velocities.



If a sweep speed for CW mode is required, bidders would need to supply a phased array probe to meet this specification; however, phased array probes are rarely included in the standard configuration of common 2D portable ultrasound systems available on the market.



Therefore, it is recommended that the requirement in Row 43 be revised as follows:



PWD sweep speed, Required"
Dalam pemeriksaan USG, fitur ini berfungsi untuk mengukur kecepatan aliran darah yang sangat tinggi secara akurat tanpa mengalami aliasing. Salah satu fungsi yang diharapkan memeriksa dan mengukur detak jantung atau fungsi katup jantung janin dalam kandungan. Fungsi ini diperlukan dalam pemeriksaan USG. Spesisikasi yang diminta tetap, tidak ada perubahan. Supply Requirements Evaluation Process Technical Evaluation