| Name Of Lots | Units | Amount |
|---|---|---|
| Lot 1: Incubator CO2 Region 1 | 165 | Rp62,370,000,000.00 |
| Lot 2: Incubator CO2 Region 2 | 167 | Rp63,126,000,000.00 |
| Lot 3: Automated Colony Counter | 16 | Rp12,574,079,999.84 |
| Published Date | Document Name |
|---|---|
| 2026-08-04 | Specific Procurement Notice - Procurement of Microbiology 2 |
| 2026-08-04 | Bidding Document - Procurement of Microbiology 2 |
| 2026-08-04 | Technical Specification - Procurement of Microbiology 2 |
| 2026-09-02 | Minutes of Prebid Meeting - Procurement of Microbiology 2 |
| 2026-09-08 | 1st Responses to Clarification of Bidding Document - Procurement of Microbiology 2 |
| No. | Date of Received | Question | Responses | Part (RFB) | Cluster | Sub Cluster |
|---|---|---|---|---|---|---|
| 1 | 2026-08-18 | 1. Row 24 – Outer Door Terkait spesifikasi “Fully insulated door with locking mechanism”, mohon klarifikasi apakah locking mechanism cukup berupa mekanisme penutup/sealing seperti gasket/karet dan inner door, atau wajib menggunakan mechanical lock atau mekanisme pengunci lainnya? |
1. Row 24 – Outer Door Regarding the "Fully insulated door with locking mechanism" specification, please clarify whether a sealing mechanism—such as a gasket/rubber seal and an inner door—is sufficient, or if a mechanical lock or other locking mechanism is mandatory. |
Supply Requirements | Evaluation Process | Technical Evaluation |
| 2 | 2026-08-18 | 2. Penyediaan CO₂ Gas Tank Terkait kewajiban penyedia menyediakan CO₂ gas tank, apakah hanya penyediaan tabung atau termasuk pengisian ulang CO₂ selama penggunaan alat? Jika termasuk, apakah biaya pengisian ulang menjadi bagian dari warranty? |
2. Provision of CO₂ Gas Tank Regarding the requirement for the provider to supply a CO₂ gas tank, does this cover only the provision of the cylinder itself, or does it include CO₂ refilling during the equipment's operation? If refilling is included, are the refill costs covered under the warranty? |
Supply Requirements | Evaluation Process | Technical Evaluation |
| 3 | 2026-08-18 | 3. Penempatan CO₂ Gas Tank Apakah CO₂ gas tank ditempatkan di dekat unit? Jika ditempatkan terpisah, mohon diinformasikan perkiraan jaraknya untuk perhitungan kebutuhan panjang selang/koneksi CO₂. atau apakah ada persyaratan minimal panjang selang/koneksinya ? Terima kasih |
3. Placement of CO₂ Gas Tank Should the CO₂ gas tank be placed near the unit? If it is to be placed separately, please provide the estimated distance so that the required length of the CO₂ hose/connection can be calculated; alternatively, are there any minimum requirements for the hose/connection length? Thank you. |
Supply Requirements | Evaluation Process | Technical Evaluation |
| 4 | 2026-08-18 | apabila perusahaan baru berdiri di tahun 2025 dan mempunyai laporan keuangan 2025, apakah diperbolehkan mengikuti pengadaan ini sebagai bidder ? | If a company is newly established in 2025 and possesses financial statements for 2025, is it permitted to participate in this procurement as a bidder? | Supply Requirements | Evaluation Process | Technical Evaluation |
| 5 | 2026-08-18 | 1. Row 24 – Outer Door Terkait spesifikasi “Fully insulated door with locking mechanism”, mohon klarifikasi apakah locking mechanism cukup berupa mekanisme penutup/sealing seperti gasket/karet dan inner door, atau wajib menggunakan mechanical lock atau mekanisme pengunci lainnya? |
1. Row 24 – Outer Door Regarding the "Fully insulated door with locking mechanism" specification, please clarify whether a sealing mechanism—such as a gasket/rubber seal and an inner door—is sufficient, or if a mechanical lock or other locking mechanism is mandatory. |
Supply Requirements | Evaluation Process | Technical Evaluation |
| 6 | 2026-08-18 | 2. Penyediaan CO₂ Gas Tank Terkait kewajiban penyedia menyediakan CO₂ gas tank, apakah hanya penyediaan tabung atau termasuk pengisian ulang CO₂ selama penggunaan alat? Jika termasuk, apakah biaya pengisian ulang menjadi bagian dari warranty? |
2. Provision of CO₂ Gas Tank Regarding the requirement for the provider to supply a CO₂ gas tank, does this cover only the provision of the cylinder itself, or does it include CO₂ refilling during the equipment's operation? If refilling is included, are the refill costs covered under the warranty? |
Supply Requirements | Evaluation Process | Technical Evaluation |
| 7 | 2026-08-18 | 3. Penempatan CO₂ Gas Tank Apakah CO₂ gas tank ditempatkan di dekat unit? Jika ditempatkan terpisah, mohon diinformasikan perkiraan jaraknya untuk perhitungan kebutuhan panjang selang/koneksi CO₂. atau apakah ada persyaratan minimal panjang selang/koneksinya ? Terima kasih |
3. Placement of CO₂ Gas Tank Should the CO₂ gas tank be placed near the unit? If it is to be placed separately, please provide the estimated distance so that the required length of the CO₂ hose/connection can be calculated; alternatively, are there any minimum requirements for the hose/connection length? Thank you. |
Supply Requirements | Evaluation Process | Technical Evaluation |
| 8 | 2026-08-18 | apabila perusahaan baru berdiri di tahun 2025 dan mempunyai laporan keuangan 2025, apakah diperbolehkan mengikuti pengadaan ini sebagai bidder ? | None | Supply Requirements | Evaluation Process | Technical Evaluation |
| 9 | 2026-08-26 | Clarification Request: Temperature Range Specification English: We refer to the tender specification requiring the CO₂ incubator temperature setting range: “At least 18°C to 50°C inclusive. Wider ranges are acceptable” (Row no. 32). We respectfully request the Tender Committee to consider revising this requirement to: Temperature setting range: “Room Temperature (RT) +3°C to 55°C. Wider ranges are acceptable.” Rationale: CO₂ incubators used for mammalian cell culture are typically designed to maintain temperatures above the ambient room temperature at 37°C. As these systems are primarily heating-controlled rather than actively refrigerated, specifying a fixed minimum temperature of 18°C may not represent a practical operating requirement under normal laboratory conditions. For example, where the laboratory ambient temperature is approximately 22°C, an incubator specified at RT +3°C would have a minimum controlled operating temperature of approximately 25°C, while comfortably covering the commonly required cell-culture temperature of 37°C. Furthermore, a range of RT +3°C to 55°C provides an extended upper operating range compared with the currently specified maximum of 50°C, while maintaining the temperature range relevant to routine CO₂ incubation applications. This revision would focus the requirement on the practical operating range and intended application of the CO₂ incubator, while allowing the Tender Committee to consider a broader range of technically suitable systems without compromising the required incubation performance. We would appreciate the Tender Committee's consideration and clarification on whether the proposed temperature range of RT +3°C to 55°C would be acceptable. Bahasa Indonesia: Kami mengacu pada spesifikasi tender yang mensyaratkan rentang pengaturan suhu inkubator CO₂: “Setidaknya 18°C hingga 50°C. Rentang yang lebih luas dapat diterima” (Baris no. 32). Kami mohon agar panitia tender mempertimbangkan untuk merevisi persyaratan ini menjadi: Rentang pengaturan suhu: “Suhu Ruangan (RT) +3°C hingga 55°C. Rentang yang lebih luas dapat diterima.” Alasan: Inkubator CO₂ yang digunakan untuk kultur sel mamalia umumnya dirancang untuk mempertahankan suhu di atas suhu ruangan, yaitu pada 37°C. Karena sistem ini terutama dikendalikan melalui pemanasan, bukan melalui pendinginan aktif, penetapan suhu minimum tetap sebesar 18°C mungkin tidak mencerminkan persyaratan operasional yang praktis dalam kondisi laboratorium normal. Sebagai contoh, jika suhu lingkungan laboratorium sekitar 22°C, inkubator yang dirancang untuk suhu RT +3°C akan memiliki suhu operasi terkendali minimum sekitar 25°C, sekaligus mampu memenuhi dengan baik suhu kultur sel yang umumnya diperlukan, yaitu 37°C. Selain itu, rentang suhu RT +3°C hingga 55°C memberikan rentang operasi atas yang lebih luas dibandingkan dengan batas maksimum yang ditetapkan saat ini, yaitu 50°C, sekaligus tetap mempertahankan rentang suhu yang relevan untuk aplikasi inkubasi CO₂ rutin. Revisi ini akan memfokuskan persyaratan pada rentang pengoperasian praktis dan tujuan penggunaan inkubator CO₂, sekaligus memungkinkan Panitia Tender untuk mempertimbangkan beragam sistem yang secara teknis sesuai tanpa mengorbankan kinerja inkubasi yang disyaratkan. Kami akan sangat menghargai pertimbangan dan klarifikasi dari Panitia Tender mengenai apakah kisaran suhu yang diusulkan, yaitu suhu kamar RT +3°C hingga 55°C, dapat diterima. |
None | Supply Requirements | Evaluation Process | Technical Evaluation |
| 10 | 2026-08-26 | Clarification Request: Temperature Range Specification Bahasa Indonesia: Kami mengacu pada spesifikasi tender yang mensyaratkan rentang pengaturan suhu inkubator CO₂: “Setidaknya 18°C hingga 50°C. Rentang yang lebih luas dapat diterima” (Baris no. 32). Kami mohon agar panitia tender mempertimbangkan untuk merevisi persyaratan ini menjadi: Rentang pengaturan suhu: “Suhu Ruangan (RT) +3°C hingga 55°C. Rentang yang lebih luas dapat diterima.” Alasan: Inkubator CO₂ yang digunakan untuk kultur sel mamalia umumnya dirancang untuk mempertahankan suhu di atas suhu ruangan, yaitu pada 37°C. Karena sistem ini terutama dikendalikan melalui pemanasan, bukan melalui pendinginan aktif, penetapan suhu minimum tetap sebesar 18°C mungkin tidak mencerminkan persyaratan operasional yang praktis dalam kondisi laboratorium normal. Sebagai contoh, jika suhu lingkungan laboratorium sekitar 22°C, inkubator yang dirancang untuk suhu RT +3°C akan memiliki suhu operasi terkendali minimum sekitar 25°C, sekaligus mampu memenuhi dengan baik suhu kultur sel yang umumnya diperlukan, yaitu 37°C. Selain itu, rentang suhu RT +3°C hingga 55°C memberikan rentang operasi atas yang lebih luas dibandingkan dengan batas maksimum yang ditetapkan saat ini, yaitu 50°C, sekaligus tetap mempertahankan rentang suhu yang relevan untuk aplikasi inkubasi CO₂ rutin. Revisi ini akan memfokuskan persyaratan pada rentang pengoperasian praktis dan tujuan penggunaan inkubator CO₂, sekaligus memungkinkan Panitia Tender untuk mempertimbangkan beragam sistem yang secara teknis sesuai tanpa mengorbankan kinerja inkubasi yang disyaratkan. Kami akan sangat menghargai pertimbangan dan klarifikasi dari Panitia Tender mengenai apakah kisaran suhu yang diusulkan, yaitu suhu kamar RT +3°C hingga 55°C, dapat diterima. |
English: We refer to the tender specification requiring the CO₂ incubator temperature setting range: “At least 18°C to 50°C inclusive. Wider ranges are acceptable” (Row no. 32). We respectfully request the Tender Committee to consider revising this requirement to: Temperature setting range: “Room Temperature (RT) +3°C to 55°C. Wider ranges are acceptable.” Rationale: CO₂ incubators used for mammalian cell culture are typically designed to maintain temperatures above the ambient room temperature at 37°C. As these systems are primarily heating-controlled rather than actively refrigerated, specifying a fixed minimum temperature of 18°C may not represent a practical operating requirement under normal laboratory conditions. For example, where the laboratory ambient temperature is approximately 22°C, an incubator specified at RT +3°C would have a minimum controlled operating temperature of approximately 25°C, while comfortably covering the commonly required cell-culture temperature of 37°C. Furthermore, a range of RT +3°C to 55°C provides an extended upper operating range compared with the currently specified maximum of 50°C, while maintaining the temperature range relevant to routine CO₂ incubation applications. This revision would focus the requirement on the practical operating range and intended application of the CO₂ incubator, while allowing the Tender Committee to consider a broader range of technically suitable systems without compromising the required incubation performance. We would appreciate the Tender Committee's consideration and clarification on whether the proposed temperature range of RT +3°C to 55°C would be acceptable. |
Supply Requirements | Evaluation Process | Technical Evaluation |
| 11 | 2026-08-28 | Izin bertanya mengenai Row 23 terkait kapasitas CO2 incubator. Row 23 menyebutkan bahwa kapasitas yang diminta hanya ukuran 100-150L yang dirasa membatasi dan hanya dimiliki oleh beberapa vendor saja seperti Thermo Scientific, Binder, dan Memmert. Faktanya banyak vendor yang memiliki CO2 incubator pada range volume 160-170L dengan dimensi yang tidak berbeda secara signifikan dibandingkan volume 150L dengan perbedaan dimensi hanya sekitar 2-5 cm. Dimensi Rata-rata Kapasitas 150L Lebar Eksterior (W): 60 – 68 cm Kedalaman / Depth (D): 70 – 78 cm Tinggi / Height (H): 85 – 90 cm Dimensi Rata-rata Kapasitas 170L Lebar Eksterior (W): 63 – 70 cm Kedalaman / Depth (D): 75 – 83 cm Tinggi / Height (H): 88 – 95 cm Apakah spesifikasi pada Row 23 bisa dibuka, agar dapat mengakomodir peserta tender secara umum? |
Regarding Row 23 on CO₂ incubator capacity: Row 23 states that the requested capacity is limited to 100–150L, which is considered restrictive and only available from a few vendors such as Thermo Scientific, Binder, and Memmert. In fact, many vendors offer CO₂ incubators in the 160–170L range with external dimensions not significantly different from the 150L models, with only about 2–5 cm variation. Average Dimensions for 150L Capacity Exterior Width (W): 60–68 cm Depth (D): 70–78 cm Height (H): 85–90 cm Average Dimensions for 170L Capacity Exterior Width (W): 63–70 cm Depth (D): 75–83 cm Height (H): 88–95 cm Could the specification in Row 23 be opened up to accommodate tender participants more broadly? |
Supply Requirements | Evaluation Process | Technical Evaluation |
| 12 | 2026-08-28 | Izin bertanya mengenai Row 23 terkait kapasitas CO2 incubator. Row 23 menyebutkan bahwa kapasitas yang diminta hanya ukuran 100-150L yang dirasa membatasi dan hanya dimiliki oleh beberapa vendor saja seperti Thermo Scientific, Binder, dan Memmert. Faktanya banyak vendor yang memiliki CO2 incubator pada range volume 160-170L dengan dimensi yang tidak berbeda secara signifikan dibandingkan volume 150L dengan perbedaan dimensi hanya sekitar 2-5 cm. Dimensi Rata-rata Kapasitas 150L Lebar Eksterior (W): 60 – 68 cm Kedalaman / Depth (D): 70 – 78 cm Tinggi / Height (H): 85 – 90 cm Dimensi Rata-rata Kapasitas 170L Lebar Eksterior (W): 63 – 70 cm Kedalaman / Depth (D): 75 – 83 cm Tinggi / Height (H): 88 – 95 cm Apakah spesifikasi pada Row 23 bisa dibuka, agar dapat mengakomodir peserta tender secara umum? |
Regarding Row 23 on CO₂ incubator capacity: Row 23 states that the requested capacity is limited to 100–150L, which is considered restrictive and only available from a few vendors such as Thermo Scientific, Binder, and Memmert. In fact, many vendors offer CO₂ incubators in the 160–170L range with external dimensions not significantly different from the 150L models, with only about 2–5 cm variation. Average Dimensions for 150L Capacity Exterior Width (W): 60–68 cm Depth (D): 70–78 cm Height (H): 85–90 cm Average Dimensions for 170L Capacity Exterior Width (W): 63–70 cm Depth (D): 75–83 cm Height (H): 88–95 cm Could the specification in Row 23 be opened up to accommodate tender participants more broadly? |
Supply Requirements | Evaluation Process | Technical Evaluation |
| 13 | 2026-08-30 | Indonesia Pertanyaan Apakah kapasitas chamber sebesar 165 L dapat diterima, sepanjang seluruh persyaratan teknis lainnya tetap terpenuhi? Penjelasan Untuk banyak laboratorium, CO₂ incubator dengan kapasitas pada rentang 100–200 L merupakan ukuran yang umum digunakan. Desain CO₂ incubator generasi terbaru dapat memberikan kapasitas chamber yang lebih besar tanpa meningkatkan ukuran eksternal alat secara signifikan. Sebagai contoh, desain terbaru dengan kapasitas 165 L dapat memiliki dimensi eksternal sekitar 637 × 900 × 880 mm, dibandingkan dengan model 150 L yang memiliki dimensi sekitar 637 × 867 × 766 mm. Dengan demikian, pengguna memperoleh tambahan sekitar 15 L ruang chamber yang dapat digunakan, namun tetap dengan footprint laboratorium yang relatif sebanding. Kapasitas tambahan tersebut juga dapat disertai dengan fitur teknologi yang lebih baru, seperti filtrasi HEPA di dalam chamber, pemulihan parameter yang lebih cepat, keseragaman temperatur yang lebih baik, user interface yang lebih advanced, serta desain pengendalian kontaminasi yang lebih baik. Oleh karena itu, kami mengusulkan agar kapasitas yang sedikit lebih besar tidak secara otomatis menggugurkan produk yang secara teknis memiliki performa dan fitur yang lebih baik. |
English Question Could a chamber capacity of 165 L be accepted, provided that all other technical requirements are met? Explanation For many laboratories, CO₂ incubators in the 100–200 L range are commonly used. Newer-generation incubator designs can provide additional usable chamber capacity without a significant increase in the external footprint. For example, a newer 165 L design may have external dimensions of approximately 637 × 900 × 880 mm, compared with a 150 L model at approximately 637 × 867 × 766 mm. This provides around 15 L of additional usable chamber space while maintaining a broadly comparable laboratory footprint. The additional capacity may also come together with newer features such as in-chamber HEPA filtration, faster parameter recovery, improved temperature uniformity, advanced user interface, and improved contamination-control design. Therefore, we suggest that slightly higher capacity should not automatically exclude an otherwise technically superior product. |
Supply Requirements | Evaluation Process | Technical Evaluation |
| 14 | 2026-08-30 | Indonesia Rentang Temperatur Pertanyaan Dapatkah Kementerian Kesehatan mengklarifikasi kebutuhan klinis atau laboratorium spesifik yang memerlukan suhu operasi minimum sebesar 18°C? Apabila operasi di bawah suhu ambient tidak merupakan kebutuhan utama dalam aplikasi yang dimaksud, apakah Kementerian Kesehatan dapat mempertimbangkan untuk merevisi persyaratan tersebut menjadi rentang operasi CO₂ incubator yang lebih umum, misalnya ambient +3°C hingga 55°C? Penjelasan Sebagian besar CO₂ incubator konvensional pada dasarnya dirancang untuk melakukan pemanasan secara terkontrol dan umumnya memiliki rentang operasi mulai dari sekitar ambient +3°C hingga 55°C. Sistem CO₂ incubator dengan performa tinggi pada kategori ini umumnya dapat memberikan: Keseragaman temperatur sekitar ±0.3°C Waktu pemulihan parameter utama sekitar 10 menit atau kurang Kelembapan relatif sekitar 93% ±2% Untuk dapat beroperasi di bawah suhu ambient, misalnya pada 15–18°C, umumnya diperlukan sistem pendinginan tambahan, seperti external chiller atau sistem pendingin berbasis Peltier yang terintegrasi. Penggunaan sistem pendinginan tersebut dapat menimbulkan beberapa trade-off teknis, antara lain: Kebutuhan sumber daya listrik tambahan Perubahan pada keseragaman temperatur hingga lebih dari 0.5°C Penurunan kelembapan relatif hingga sekitar 4% Untuk aplikasi rutin CO₂ incubation dan cell culture, kestabilan dan keseragaman temperatur, pemulihan parameter yang cepat, serta kelembapan relatif yang tinggi merupakan parameter performa yang sangat penting. Oleh karena itu, apabila tidak terdapat aplikasi spesifik yang benar-benar membutuhkan operasi pada suhu 18°C, kami mengusulkan agar batas minimum temperatur dipertimbangkan berdasarkan suhu ambient. |
English Temperature Range Question Could MoH clarify the specific clinical or laboratory application that requires a minimum operating temperature of 18°C? If below-ambient operation is not essential to the intended application, could MoH consider revising the requirement to a more conventional CO₂ incubator operating range, such as ambient +3°C to 55°C? Explanation Most conventional CO₂ incubators are designed primarily for controlled heating and typically operate from approximately ambient +3°C to 55°C. High-performance systems in this category can typically maintain: Temperature uniformity of around ±0.3°C Recovery of key parameters within approximately 10 minutes or less Relative humidity of around 93% ±2% To operate below ambient temperature, for example at 15–18°C, an additional cooling system is generally required, such as an external chiller or an integrated Peltier-based cooling system. These cooling systems may introduce technical trade-offs, including: Additional power supply requirements Changes in temperature uniformity of more than 0.5°C Reduction in relative humidity by up to approximately 4% For routine CO₂ incubation and cell-culture applications, stable temperature uniformity, rapid recovery, and high relative humidity are generally critical performance parameters. Therefore, unless there is a specific application requiring operation at 18°C, we suggest considering an ambient-relative minimum temperature instead. |
Supply Requirements | Evaluation Process | Technical Evaluation |
| 15 | 2026-08-30 | Indonesia Rentang Temperatur Pertanyaan Dapatkah Kementerian Kesehatan mengklarifikasi kebutuhan klinis atau laboratorium spesifik yang memerlukan suhu operasi minimum sebesar 18°C? Apabila operasi di bawah suhu ambient tidak merupakan kebutuhan utama dalam aplikasi yang dimaksud, apakah Kementerian Kesehatan dapat mempertimbangkan untuk merevisi persyaratan tersebut menjadi rentang operasi CO₂ incubator yang lebih umum, misalnya ambient +3°C hingga 55°C? Penjelasan Sebagian besar CO₂ incubator konvensional pada dasarnya dirancang untuk melakukan pemanasan secara terkontrol dan umumnya memiliki rentang operasi mulai dari sekitar ambient +3°C hingga 55°C. Sistem CO₂ incubator dengan performa tinggi pada kategori ini umumnya dapat memberikan: Keseragaman temperatur sekitar ±0.3°C Waktu pemulihan parameter utama sekitar 10 menit atau kurang Kelembapan relatif sekitar 93% ±2% Untuk dapat beroperasi di bawah suhu ambient, misalnya pada 15–18°C, umumnya diperlukan sistem pendinginan tambahan, seperti external chiller atau sistem pendingin berbasis Peltier yang terintegrasi. Penggunaan sistem pendinginan tersebut dapat menimbulkan beberapa trade-off teknis, antara lain: Kebutuhan sumber daya listrik tambahan Perubahan pada keseragaman temperatur hingga lebih dari 0.5°C Penurunan kelembapan relatif hingga sekitar 4% Untuk aplikasi rutin CO₂ incubation dan cell culture, kestabilan dan keseragaman temperatur, pemulihan parameter yang cepat, serta kelembapan relatif yang tinggi merupakan parameter performa yang sangat penting. Oleh karena itu, apabila tidak terdapat aplikasi spesifik yang benar-benar membutuhkan operasi pada suhu 18°C, kami mengusulkan agar batas minimum temperatur dipertimbangkan berdasarkan suhu ambient. |
English Temperature Range Question Could MoH clarify the specific clinical or laboratory application that requires a minimum operating temperature of 18°C? If below-ambient operation is not essential to the intended application, could MoH consider revising the requirement to a more conventional CO₂ incubator operating range, such as ambient +3°C to 55°C? Explanation Most conventional CO₂ incubators are designed primarily for controlled heating and typically operate from approximately ambient +3°C to 55°C. High-performance systems in this category can typically maintain: Temperature uniformity of around ±0.3°C Recovery of key parameters within approximately 10 minutes or less Relative humidity of around 93% ±2% To operate below ambient temperature, for example at 15–18°C, an additional cooling system is generally required, such as an external chiller or an integrated Peltier-based cooling system. These cooling systems may introduce technical trade-offs, including: Additional power supply requirements Changes in temperature uniformity of more than 0.5°C Reduction in relative humidity by up to approximately 4% For routine CO₂ incubation and cell-culture applications, stable temperature uniformity, rapid recovery, and high relative humidity are generally critical performance parameters. Therefore, unless there is a specific application requiring operation at 18°C, we suggest considering an ambient-relative minimum temperature instead. |
Supply Requirements | Evaluation Process | Technical Evaluation |
| 16 | 2026-08-30 | Indonesia Pertanyaan Apakah kapasitas chamber sebesar 165 L dapat diterima, sepanjang seluruh persyaratan teknis lainnya tetap terpenuhi? Penjelasan Untuk banyak laboratorium, CO₂ incubator dengan kapasitas pada rentang 100–200 L merupakan ukuran yang umum digunakan. Desain CO₂ incubator generasi terbaru dapat memberikan kapasitas chamber yang lebih besar tanpa meningkatkan ukuran eksternal alat secara signifikan. Sebagai contoh, desain terbaru dengan kapasitas 165 L dapat memiliki dimensi eksternal sekitar 637 × 900 × 880 mm, dibandingkan dengan model 150 L yang memiliki dimensi sekitar 637 × 867 × 766 mm. Dengan demikian, pengguna memperoleh tambahan sekitar 15 L ruang chamber yang dapat digunakan, namun tetap dengan footprint laboratorium yang relatif sebanding. Kapasitas tambahan tersebut juga dapat disertai dengan fitur teknologi yang lebih baru, seperti filtrasi HEPA di dalam chamber, pemulihan parameter yang lebih cepat, keseragaman temperatur yang lebih baik, user interface yang lebih advanced, serta desain pengendalian kontaminasi yang lebih baik. Oleh karena itu, kami mengusulkan agar kapasitas yang sedikit lebih besar tidak secara otomatis menggugurkan produk yang secara teknis memiliki performa dan fitur yang lebih baik. |
English Question Could a chamber capacity of 165 L be accepted, provided that all other technical requirements are met? Explanation For many laboratories, CO₂ incubators in the 100–200 L range are commonly used. Newer-generation incubator designs can provide additional usable chamber capacity without a significant increase in the external footprint. For example, a newer 165 L design may have external dimensions of approximately 637 × 900 × 880 mm, compared with a 150 L model at approximately 637 × 867 × 766 mm. This provides around 15 L of additional usable chamber space while maintaining a broadly comparable laboratory footprint. The additional capacity may also come together with newer features such as in-chamber HEPA filtration, faster parameter recovery, improved temperature uniformity, advanced user interface, and improved contamination-control design. Therefore, we suggest that slightly higher capacity should not automatically exclude an otherwise technically superior product. |
Supply Requirements | Evaluation Process | Technical Evaluation |